User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Honeywell UDC 4000:Digital Controller øF DATE: 06-May-96 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 14 NUMBER OF LINES: 127 CONFIGURATION: Fluke 5700A STANDARD: Omega TRC III ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON # This procedure is run with the MET/CAL Configuration #1 loaded. 1.001 ASK- P F W 1.002 ASK+ X 1.003 HEAD ÍÍINITIAL CONDITIONS ÍÍ 1.004 STD Omega TRC III 1.005 DISP The following additional equipment is required: 1.005 DISP Omega TRC III Ice Point Cell 1.005 DISP Omega Type "T" Reference Thermocouple 1.006 HEAD ÍÍ INITIAL CONDITIONS ÍÍ 1.007 DISP Insert the reference thermocouple into the center 1.007 DISP well of the Ice Point Cell. 1.008 DISP Connect the copper end (white connector) to the 1.008 DISP 5700 OUTPUT HI/LO. Red wire is HI, green wire is LO. 1.009 DISP Connect a "T" type connector with "T" type T/C wire 1.009 DISP to UUT as follows: 1.009 DISP ÿ [27][91]1m "T" probe TO UDC 4000 1.009 DISP ÿ OUTPUT + ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ terminal #39 1.009 DISP ÿ OUTPUT - ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ terminal #40 1.010 DISP Connect the UUT to AC power as follows: 1.010 DISP ÿ AC HOT ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ terminal #3 1.010 DISP ÿ AC NEUTRAL ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ terminal #2 1.011 DISP Allow the UUT a 1 hour warm-up. 1.012 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.013 HEAD {º ThermoCouple Tests º} 1.014 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.015 HEAD {ÍÍ "T" Type T/C ÍÍ} 1.016 JMP 2.001 1.017 EVAL 2.001 5700 -5.341mV S 2W 2.002 MEMI Use the keyboard to enter the UUT reading in øF. 2.003 ACC -300.0øF 0.10U 2.004 MEME 2.005 MEMC øF 2.0U #! Test Tol 2, Sys Tol 0.1, TUR 20.000 (>= 4.00). 3.001 5700 -4.149mV S 2W 3.002 MEMI Use the keyboard to enter the UUT reading in øF. 3.003 ACC -200.0øF 0.10U 3.004 MEME 3.005 MEMC øF 2.0U #! Test Tol 2, Sys Tol 0.1, TUR 20.000 (>= 4.00). 4.001 5700 -2.581mV S 2W 4.002 MEMI Use the keyboard to enter the UUT reading in øF. 4.003 ACC -100.0øF 0.10U 4.004 MEME 4.005 MEMC øF 2.0U #! Test Tol 2, Sys Tol 0.1, TUR 20.000 (>= 4.00). 5.001 5700 -0.674mV S 2W 5.002 MEMI Use the keyboard to enter the UUT reading in øF. 5.003 ACC 0.0øF 0.10U 5.004 MEME 5.005 MEMC øF 2.0U #! Test Tol 2, Sys Tol 0.1, TUR 20.000 (>= 4.00). 6.001 5700 0.000mV S 2W 6.002 MEMI Use the keyboard to enter the UUT reading in øF. 6.003 ACC 32.0øF 0.1U 6.004 MEME 6.005 MEMC øF 2.0U #! Test Tol 2, Sys Tol 0.1, TUR 20.000 (>= 4.00). 7.001 5700 1.518mV S 2W 7.002 MEMI Use the keyboard to enter the UUT reading in øF. 7.003 ACC 100.0øF 0.1U 7.004 MEME 7.005 MEMC øF 2.0U #! Test Tol 2, Sys Tol 0.1, TUR 20.000 (>= 4.00). 8.001 5700 3.967mV S 2W 8.002 MEMI Use the keyboard to enter the UUT reading in øF. 8.003 ACC 200.0øF 0.1U 8.004 MEME 8.005 MEMC øF 2.0U #! Test Tol 2, Sys Tol 0.1, TUR 20.000 (>= 4.00). 9.001 5700 6.647mV S 2W 9.002 MEMI Use the keyboard to enter the UUT reading in øF. 9.003 ACC 300.0øF 0.1U 9.004 MEME 9.005 MEMC øF 2.0U #! Test Tol 2, Sys Tol 0.1, TUR 20.000 (>= 4.00). 10.001 5700 9.523mV S 2W 10.002 MEMI Use the keyboard to enter the UUT reading in øF. 10.003 ACC 400.0øF 0.1U 10.004 MEME 10.005 MEMC øF 2.0U #! Test Tol 2, Sys Tol 0.1, TUR 20.000 (>= 4.00). 11.001 5700 12.572mV S 2W 11.002 MEMI Use the keyboard to enter the UUT reading in øF. 11.003 ACC 500.0øF 0.1U 11.004 MEME 11.005 MEMC øF 2.0U #! Test Tol 2, Sys Tol 0.1, TUR 20.000 (>= 4.00). 12.001 5700 15.769mV S 2W 12.002 MEMI Use the keyboard to enter the UUT reading in øF. 12.003 ACC 600.0øF 0.1U 12.004 MEME 12.005 MEMC øF 2.0U #! Test Tol 2, Sys Tol 0.1, TUR 20.000 (>= 4.00). 13.001 ACC 19.095mV 0.0034U 13.002 MESS Adjust stimulus so the UUT just changes from 13.002 MESS 700øF to 701øF. 13.003 5700 19.095mV S 2W 13.004 MEMI Use the keyboard to enter the UUT reading in øF. 13.005 ACC 700.0øF 0.1U 13.006 MEME 13.007 MEMC øF 2.0U #! Test Tol 2, Sys Tol 0.1, TUR 20.000 (>= 4.00). 14.001 END 14.002 HEAD {} 14.003 DISP {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 14.003 DISP {º THIS COMPLETES THE VERIFICATION º} 14.003 DISP {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 14.004 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.